Search for Products

Ismatec pumps and tubing



Swiss designed multichannel pumps with a quality build and the best UK prices.

IPC and IPC-N

> Operating Manual (PDF)
..
Model Order No. Channels Flow rates
ml/ min
per channel
Dimensions Weight
D x W x H (mm) D x W x H (inches) kg lb
IPC 4 ISM 930 4 0.002 – 44 180 x 175 x 130 7 1/16 x 6 7/8 x 5 1/8 4.6 10
IPC 8 ISM 931 8   220 x 175 x 130 8 2/3 x 6 7/8 x 5 1/8 5.1 12
IPC 12 ISM 932 12 260 x 175 x 130 10¼ x 6 7/8 x 5 1/8 5.8 13
IPC 16 ISM 933 16 300 x 175 x 130 11 7/8 x 6 7/8 x 5 1/8 6.5 15
IPC 24 ISM 934 24 380 x 175 x 130 15 x 6 7/8 x 5 1/8 7.9 18

IPC-N 4ISM 93540.0004 – 11180 x 175 x 1307 1/16 x 6 7/8 x 5 1/84.610
IPC-N 8ISM 9368 220 x 175 x 1308 2/3 x 6 7/8 x 5 1/85.112
IPC-N 12ISM 93712260 x 175 x 13010¼ x 6 7/8 x 5 1/85.813
IPC-N 16ISM 93816300 x 175 x 13011 7/8 x 6 7/8 x 5 1/86.515
IPC-N 24ISM 93924380 x 175 x 13015 x 6 7/8 x 5 1/87.918
 
4 channels
IPC 4

8 channels
IPC 8

12 channels
IPC 12

16 channels
IPC 16

24 channels
IPC 24
Specifications:
Characteristics - microprocessor controlled
- planetary drive with 8 actively driven stainless steel rollers
- low pulsation
- high repeatability on all channels
- paint finished stainless steel housings
Tubing type 2-stop color-coded tubing, 0.13 3.17 mm i.d.
Tube-bed - CA Click'n'go cassettes
- simple handling
- automatic occlusion setting
- defined and reproducible occlusion conditions
- each channel allows individual tubing selection
Operating panel - 6-button membrane key-pad
- LED display
- MAX key for fast filling or emptying the tubing system
Operating modes - pumping by drive speed (in %)
- pumping by flow rate (ml/min)
- factory-set flow rates for all available tube sizes
- dispensing by volume (ml)
- dispensing by time (0.1s up to 999hrs)
- dispensing by volume within a pre-set time
- interval dispensing by volume with a pause
- interval dispensing by time with a pause
- setting the number of dispensing cycles
- pause time setting (from 0.1s up to 999hrs)
- roller back-steps for drip-free dispensing

- stand-by mode — prevents the tubing system from drying out
- reversible rotation direction
Calibrating function - for flow rate (ml/min)
- for volume (ml)
Rotor speed - 0.4 45.0 rpm (IPC)
- 0.11 11.25 rpm (IPC-N)
Speed setting - in %, resolution 0.1% (rpm)
- in µl/min or ml/min (flow rate)
Speed control closed loop control for load independent speed
RS232 interface RS232 "in" - for operation controlling via PC
RS232 "out" - for cascading up to 8 pumps
Digital inputs (TTL level) Run/Stop, rotation direction, autostart
Analog input Speed control (05 or 010V / 020 or 420mA)
Analog output Speed monitoring (010VDC)
Differential pressure 1.0 bar (14.5 psi) possible (depending on tubing material);
small tubing diameters and/or cassettes with pressure lever allow higher pressures
Motor type DC motor
Power consumption   30 W
Power supply 115/230VAC; 50/60Hz
Protection rating IP 30
Dimensions/Weight see table above
Accessories - cassettes
- foot switch

- Driver for "LabView" software (free download)
 
   

Overview IP/IPC series